Lars-Erik Briggner, Assignment 3
1. Calculate the intensity values for 200- and 100 reflection for NaCl.
i) dhkl = a/(h2+k2+l2)½,, i.e. d200 = 5.638/2 = 2.819 Å. l=2dsinq i.e. sinq/l=(2d)-1 = (2x2.819)-1= 0.177Å
From the Int. Cryst. Tables can scattering factors for 0.177 be interpolated: fNa= 8.69, fCl= 13.08
Comment: Correct.
5 Marks
Using the Structure Factor equation: F = Sfne2pi(hx+ky+lz) gives F2002 = (cosine part)2 + (sine part)2
Cosine part = Sfncos{cos(hx+ky+lz)}= 8.69cos{360(2x0+0x0+0x0)}+....according the example +
13.08cos{360(2x½+0x0+0x0)}+.....= 87.08, likewise the Sine part = 0
F2002= I200= 87.082= 7583 units
Comment: Correct.
5 Marks
ii) I100=?. d100=5.638/1=5.638 Å. sinq/l=(2x5..638)-1=0.089. The scatter factors become irrelevant since
in the Structure Factor equation both the Cosine and the Sine part become zero which means that I100=0 units
The reflection does not exist.
Comment: You are correct that the intensity is zero,
meaning in effect that the reflection does not exist. The reason for this is
the symmetry of the unit cell with cancelling of contributions from the various
ions; this is explained fully in the model answer.
3 Marks
2. What is the multplicity for the 200 reflection?
2,0,0; -2,0,0; 2,-0,0; -2,-0,0; 2,0,-0; -2,0,-0; 2,-0,-0; -2,-0,-0 but since 0 and -0 is the same there are only
two differents sets of planes, thus J200=2
Comment: The logic is correct as far as
putting h=2 or h=-2; however being cubic, the other two axes are
identical so 020, 0-20, 002, 00-2 are also permissable giving a final
multiplicity of 6.
1 Mark
3. What is the polarisation allowance factor the 200 reflection?
d200 = 2.819 Å. Ka(Cu)=1.54 Å, i.e. 1.54=2x2.819sinq, thus q=15.85° and 2q=31.70°
P=(1+cos²2q)/2= 0.862, i.e. the intensity will be reduced to about 86%
Polarisation produces a 86.2% reduction in intensity for the 200 reflection.
Comment: Correct.
4 Marks
4. What is the lorentz allowance factor neede for the 200 reflection?
L= 1/(sin²qcosq), q=15.85 i.e. L = 13.94
Comment: Correct.
3 Marks
5. What is the intensity ratio I111/I200?
I111/I200= (J111P111L111F111²)/(J200P200L200F200²), I200= 182239, I111=52347, i.e. approx 0.29
Comment: The method is correct but of course you
are carrying a wrong value (2) for the multiplicity (6) so the actual
numerical result obtained (0.288) is too large by a factor of 3.
3 Marks
6. List plausible factors that could influence measured I111/I200 ratio!
i) Preferred orientation - one reflection will have higher probability than the other due to orientation of crystallites
ii) Texture - To few and large particles to give a good statistical base
iii) Spinning - No spinning means that not the whole sample will be averaged, biased
iv) Sampling time - The sampling time should be adjusted following the wavelength in order to compensate.
v) Temperature - changes in temperature will affect different reflection to a different extent
vi) X-ray absorption - ....
vii) Transparency - different reflections will be affected to different extent especially with thick organic samples
viii) Primary slit - The primary slit should be adjusted following different 2q
ix) Monochromator - bad adjustment could affect the reflections to a different extent
x) Filter - same as above
xi) Alignment - The instrumet need optimum alignment in order to perform with accuracy
xii) Detector type - A bad choice of detector could produce erroneous results
xiii) A badly stabilised power supply -.....
Comment: You have correctly identified several
prime suspects for erroneous intensity ratios (your numbers i, ii, iii, xi).
The others are also plausible though I might make the following caveats:
v) temperature changes might upset the ratios provided the changes are such that
one peak is measured consitently at a different temperature to other peak(s);
vii) absorption can indeed cause problems in principle but note this is an
NaCl sample, not an organic;
xii) the detector similarly might upset ratios if the detector is unstable with
time (and therefore operating differently for different reflections) or in the
case of a position sensitive detector it is non-linear with respect to position.
In the case of the monochromator/filter (ix, x) this is less likely to cause a
ratio upset since any such fault is likely to affect all peaks equally, and
therefore the ratio will be unaffected.
8 Marks overall for this question