Problem Set 3 - Wolfgang Seidl

1 a) Calculated intensity from the structure factor equation, for the 200 reflection:

        d-spacing of the 200 reflection (cubic system):

            dhkl = a/Ö(h2 + k2 + l2 ); a = 5.638 Å

            d200 = 5.638/Ö(22 + 02 + 02) = 5.638/Ö4 = 2.819 Å

        From Bragg´s Law:

           sin q/l = 1/2d = 1/(2 * 2.819Å) = 0.1774

        Finally from ICT by interpolation, assuming that there is a linear decrease, the calculated factors are:

            D (fNa+) = 0.645 and D (fCl-) = 1.92 for sin q/l between 0.15 and 0.20.

             fNa+ = 9.035 - [D (fNa+)*(0.1774-0.15)/0.05] = 8.682

            fCl- = 14.12 - [D (fCl-)*(0.1774-0.15)/0.05] = 13.068

        cosine part of  F200:

        S fn cos{360 (hx + ky + lz)}=

        =   8.682 cos{360(2*0 + 0*0 + 0*0)}      +   8.682 cos{360(2*0 + 0*0.5 + 0*0.5)}+
        +   8.682 cos{360(2*0.5 + 0*0 + 0*0.5)}+   8.682 cos{360(2*0.5 + 0*0.5 + 0*0)}+
        + 13.068 cos{360(2*0.5 + 0*0 + 0*0)}   + 13.068 cos{360(2*0 + 0*0.5 + 0*0)}   +
        + 13.068 cos{360(2*0 + 0*0 + 0*0.5)}   + 13.068 cos{360(2*0.5 + 0*0.5 + 0*0.5)} =

        = 8.682 (cos 0 + cos 0 + cos 360 + cos 360) + 13.068 (cos 360 + cos 0 + cos 0 + cos 360) = 4 * 8.682 + 4 * 13.068 =
        = 87

        sine part of F200:

        S fn sin{360 (hx + ky + lz)}=

         =   8.682 sin{360(2*0 + 0*0 + 0*0)}     +   8.682 sin{360(2*0 + 0*0.5 + 0*0.5)}+
        +   8.682 sin{360(2*0.5 + 0*0 + 0*0.5)}+   8.682 sin{360(2*0.5 + 0*0.5 + 0*0)}+
        + 13.068 sin{360(2*0.5 + 0*0 + 0*0)}   + 13.068 sin{360(2*0 + 0*0.5 + 0*0)}   +
        + 13.068 sin{360(2*0 + 0*0 + 0*0.5)}   + 13.068 sin{360(2*0.5 + 0*0.5 + 0*0.5)} =

        = 8.682 (sin 0 + sin 0 + sin 360 + sin 360) + 13.068 (sin 360 + sin 0 + sin 0 + sin 360) = 8.682 * 0 + 13.068 * 0 =
        = 0

After combining the two parts one gets:

        F200 = 87 + i * 0, with F2002 = I200 =  872 + 02 = 7569 units


1 b) Calculated intensity from the structure factor equation, for the 100 reflection:

        d-spacing of the 200 reflection (cubic system):

            dhkl = a/Ö(h2 + k2 + l2 ); a = 5.638 Å

            d100 = 5.638/Ö(12 + 02 + 02) = 5.638/Ö1 = 5.638 Å

        From Bragg´s Law:

            sin q/l = 1/2d = 1/(2 * 5.638Å) = 0.0886

        As in 1 a) one would have to search the values above and below sin q/l for the 100 reflection and calculate the correct value by linear interpolation.
        Knowing that there is something strange about this "reflection" I did the calculation for the structure factor equation first with the result that it is 0!
        Hence I didn´t look up the scattering factors for Na+ and Cl-.

        cosine part of  F100:

        S fn cos{360 (hx + ky + lz)}=

        =   8.682 cos{360(1*0 + 0*0 + 0*0)}      +   8.682 cos{360(1*0 + 0*0.5 + 0*0.5)}+
        +   8.682 cos{360(1*0.5 + 0*0 + 0*0.5)}+   8.682 cos{360(1*0.5 + 0*0.5 + 0*0)}+
        + 13.068 cos{360(1*0.5 + 0*0 + 0*0)}   + 13.068 cos{360(1*0 + 0*0.5 + 0*0)}   +
        + 13.068 cos{360(1*0 + 0*0 + 0*0.5)}   + 13.068 cos{360(1*0.5 + 0*0.5 + 0*0.5)} =

        = 8.682 (cos 0 + cos 0 + cos 180 + cos 180) + 13.068 (cos 180 + cos 0 + cos 0 + cos 180) = 8.682 * (1+1-1-1) + XXX * (-1+1+1-1) =
        = 0

        sine part of F200:

        S fn sin{360 (hx + ky + lz)}=

         =   8.682 sin{360(1*0 + 0*0 + 0*0)}     +   8.682 sin{360(1*0 + 0*0.5 + 0*0.5)}+
        +   8.682 sin{360(1*0.5 + 0*0 + 0*0.5)}+   8.682 sin{360(1*0.5 + 0*0.5 + 0*0)}+
        + 13.068 sin{360(1*0.5 + 0*0 + 0*0)}   + 13.068 sin{360(1*0 + 0*0.5 + 0*0)}   +
        + 13.068 sin{360(1*0 + 0*0 + 0*0.5)}   + 13.068 sin{360(1*0.5 + 0*0.5 + 0*0.5)} =

        = 8.682 (sin 0 + sin 0 + sin 180 + sin 180) + 13.068 (sin 180 + sin 0 + sin 0 + sin 180) = 8.682 * 0 + 13.068 * 0 =
        = 0

After combining the two parts one gets:

        F100 = 0 + i * 0, with F1002 = I100 =  02 + 02 = 0 units

This result is not astonishing, because in cubic systems there is high symmetry which leads to certain (symmetry-specific) exclusions. 100 is such an exclusion in cubic systems. Also 010 and 001 would give no reflection. The unit cell defines all atom places which are introduced later into the structure factor equation. Due to the fact that the sine and cosine part are made up of sums it is possible under certain circumstances that both parts of the structure factor equation can become zero. Result of this is that such a specific reflection shows no intensity and hence cannot be measured..
In opposite of this an elongation of one axis would allow to observe I100. In a triclinic or orthorhombic unit cell all three possible reflections 100, 010 and 001 could be measured. Each of them then belongs to one of the three axises.


2 ) Multiplicity for 200 reflection:

        2,0,0; -2,0,0; 2,-0,0; -2,-0,0; 2,0,-0; -2,0,-0; 2,-0,-0; -2, -0, -0.

        Knowing that -0 and +0 are the same there remain only two senseful and logical permutations: 200 and -200!
        Hence j = 2 (in opposite of j = 8 for the 111 reflection).

        For 100 j would be the same as for 200, j = 2

3 ) Polarisation for 200 reflection:

       d200 = 2.819 Å, l = 1.54 Å are put into the Bragg equation which gives sin q = 0.273. Hence q = 15.85 ° and 2 q = 31.7 °. 4 ) Lorentz Factor for 200 reflection:

        q = 15.85 °; sin q = 0.273; sin2 q = 0.0746 and cos q = 0.962.

        L = 1/sin2 q cos q = 13.936

5 ) Intensity ratio I111/I200:

        I = j * P * L * A * F2, with a neglible value for A if it is a flat sample and Bragg Brentano geometry

        I111 = 8 * 0.894 * 18.39 * 398 = 52347 c units
        I200 = 2 * 0.862 * 13.936 * 7563 = 181706 c units

         I111 /I200 = 0.288; I200 is approximately 3,5 times stronger than I111

6 ) Reasons for a different intensity ratio in comparison to the experiment:
 

Sample treatment:
How is the sample prepared? Is it sieved, or is it just ground in a mortar or mill?
The preparation might lead to preferred orientation or texture effects in the sample. If one of the two cases occurs the intensity information can be seriously influenced! A comparison of different samples of the same material might give different intensities!

Spinning or not?
If the sample is measured without spinning it it migh lead to a preferred orientation, which increases the intensity for reflections of the preferred orientation.

Absorption correction:
Depending on how the sample is measured a specific absorption correction for transmission or reflection geometry has to be made!

Temperature effects:
Temperature has an influence on the excursions of atoms in a crystal. Hence intensity information can be altered by measurements at different temperatures. A cooled sample would have other intensities as a heated one.

Badly stabilised power supply for the X-ray tube:
This might lead to a change in intensity of the original emitted radiation of the X-ray tube.

Badly stabilised power supply for the detector:
If the high voltage supply of a scintillation detector is not stable, the multiplication of an incoming signal would change and lead to different measured intensities for the same signal! A PSD detector loses the linearity with high count rates. Therefore high intensities might not be as exact as low ones.

Polarisation of the monochromator:
A monochromator polarises the X-ray beam and hence the intensity of the diffraction pattern. The absorption correction should be small and mostly neglible, but two different monochromators (same type) might give slighlty different intensity informations in the same measurement setup.

Scattering factors:
It is assumed that there is a linear change in the scattering factor in the ICT. How good is this assumption? High intensities are more influenced by a slight change in the scattering factors due to the product which is formed for the intensity.