Given parameters
I(l) = 10
photons/sec
I0(l)= 108
photons/sec
(m/r)1.54 = 230 cm2/g
r = 11.3 g/cc
Substituting
these values in the equation,
I(l) / Io(l) = exp{- (m / r)l rx}
The final value obtained was
53 mm.
The much higher thickness than the above value for a beam stop in an unfiltered non-monochromated copper X-ray source could be due to absorb the high energy background radiation.
2. Powder diffraction patterns of organic compounds are often
broadened due to sample transparency. Briefly discuss how
narrow peaks might be obtained from a laboratory diffractometer configured for
Bragg-Brentano geometry. Briefly explain how and why narrow peaks can be
obtained with organic samples on synchrotron angle-dispersive powder
diffractometers.
The best way to obtain the narrow peak in
Bragg-Brentano geometry is to grind the sample into fine powder. Other
alternative way, for organic compounds in particular, is to apply a fine
coating of the powder on a Silicon single crystal.
However, the diffraction pattern produced by
synchroton radiation is much narrower due to the two important properties of
such radiations. Firstly, the synchroton radiation is highly intense leading to
better signal:noise ratio. Secondly,
the near parallel beam obtained with the synchroton radiations eliminates the
low angle peak broadening found in conventional laboratory diffractometer.
3.Crystal
monochromators exploit the fact that only planes in the Bragg
condition diffract X-rays in the required direction. On synchrotron powder
diffractometers, double-bounce monochromators are rotated about
a 2q axis so as to tune the instrument to a
particular wavelength. If the wavelength on the instrument has to be changed
from 0.75 to 0.65 Å (e.g. to avoid the high absorption of a zirconium
compound), by what angle would the monochromator have to be rotated if
Si (111) planes are used? (The d-spacing for Si (111) planes is 3.136 Å.) Show your
workings clearly in your answer.
Given parameters
l = 0.65 Å
d = 3.136 Å
Substituting these values in
the equation l = 2d sinq one can calculate q. From this the 2q can be calculated and found to
be 11.9°. Thus the double bounce monochromator is to be rotated
about an able 11.9° in order to use the Si (111) plane.
E = 12.398 / l
What range of d-spacings
can be measured for a detector designed to record X-ray photons with energies
in the range 20-100 keV when the detector angle (2q) is set to 2.7°. Show your workings
clearly in your answer.
Given
parameter,
Energy
= 20 – 100 keV
2q = 2.7°
Using
the relation
Ed
sin q = C
Where
C= 6.1993 keV Å. Substituting values of
20 keV and 100 keV in the above equation and 1.35 for q, the range of d-spacings
have been found to be ranging from 2.631Å
to 13.156 Å.
Ans: