GENERAL COMMENT: This particular submission gave us
much grief since it failed to display properly on non Microsoft-based
computers. On this occasion JKC edited the HTML so that it strictly
conformed to the HTML taught in the course material. For future submissions,
you should not use products like MS-Word 2000. Note that the project work
is partly judged on the quality of the HTML that you write. Therefore you should
use the assignments (which do not give marks for HTML except for the first
assignment set) as practice for writing HTML.
Please compare the HTML as modified by JKC to that produced by Word and look
at the relative simplicity of the marked version (which could be
made even simpler)
Given parameters
I(l) = 10 photons/sec
I0(l)= 108 photons/sec
(m/r)1.54 = 230 cm2/g
r = 11.3 g/cc
Substituting these values in the equation,
I(l) / Io(l) = exp{- (m / r)l rx}
The final value obtained was 53 mm.
HTML comment: when viewed with my browser the value
originally appeared to be 53 mm (i.e. millimetres!).
Hence my comments about your source HTML
Comment: You took the log of 10-8 by mistake, and
you were asked to give the answer in mm and not microns. It would be better
style to show your workings in more detail
1 Mark
The much higher thickness than the above value for a beam stop in an unfiltered non-monochromated copper X-ray source could be due to absorb the high energy background radiation.
Comment: correct but a bit too brief
1 Mark
2.
Powder diffraction patterns of organic compounds are often
broadened due to sample
transparency. Briefly discuss how
narrow peaks might be obtained from a laboratory diffractometer configured for
Bragg-Brentano geometry. Briefly explain how and why narrow peaks can be
obtained with organic samples on synchrotron angle-dispersive powder
diffractometers.
The best way to obtain the narrow peak in Bragg-Brentano geometry is to grind the sample into fine powder. Other alternative way, for organic compounds in particular, is to apply a fine coating of the powder on a Silicon single crystal.
Comment: only partially correct; see model answer
1 Mark
However, the diffraction pattern produced by synchroton radiation is much narrower due to the two important properties of such radiations. Firstly, the synchroton radiation is highly intense leading to better signal:noise ratio. Secondly, the near parallel beam obtained with the synchroton radiations eliminates the low angle peak broadening found in conventional laboratory diffractometer.
Comment: only partially correct; see model answer
1 Mark
3.Crystal
monochromators exploit the fact that only planes in the Bragg
condition diffract X-rays in the required direction. On synchrotron powder
diffractometers,
double-bounce monochromators are rotated about
a 2q
axis so as to tune the instrument to a
particular wavelength. If the wavelength on the instrument has to be changed
from 0.75 to 0.65 Å (e.g. to avoid the high absorption of a zirconium
compound), by what angle would the monochromator have to be rotated if
Si (111) planes are used? (The d-spacing for
Si (111) planes is 3.136 Å.) Show your
workings clearly in your answer.
Given parameters
l = 0.65 Å
d = 3.136 Å
Substituting these values in the equation l = 2d sinq one can calculate q. From this the 2q can be calculated and found to be 11.9°. Thus the double bounce monochromator is to be rotated about an able 11.9° in order to use the Si (111) plane.
Comment: you have missed the point of the question
since it is a change in angle that was to be calculated. See model answer
for further details
1 Mark
E = 12.398 / l
What range of d-spacings
can be measured for a detector designed to record X-ray photons with energies
in the range 20-100 keV when the detector angle
(2q) is set to 2.7°. Show your workings
clearly in your answer.
Given parameter,
Energy = 20 - 100 keV
2q = 2.7°
Using the relation
Ed sin q = C
Where C= 6.1993 keV Å. Substituting values of 20 keV and 100 keV in the above equation and 1.35 for q, the range of d-spacings have been found to be ranging from 2.631Å to 13.156 Å.
Comment: the question specifically asked
you to show your workings clearly -
you risk losing many marks in the event of a small numerical error
4 Marks
Ans: